            文章詳情

            A10CD電導率測試儀故障判斷及排除

            日期:2025-07-02 01:02
            瀏覽次數:434
            摘要:A10CD電導率測試儀故障判斷及排除

            A10CD電導率測試儀故障判斷及排除

            現象 可能因素 排除方法
            1.儀表無顯示
            A.電源沒接通

            B.儀表故障
            A.檢查(01)(03)之間有無220V電壓
            B.請專業人員維修
            2.顯示不穩定
            A.電極接線有誤
            B.管路中有氣泡
            C.水質不穩定
            D.電源有強干擾
            A.檢查電纜線有無接錯
            B.整改管路或另選測量點
            C.用穩定水源排除儀表原因
            D.視原因對電源采取措施
            3.讀數誤差大
            A.常數設置有誤
            B.電極常數發生改變
            C.測點流速太急或死水
            A.重新設置常數
            B.更換新電極
            C.將電極安裝于流速較緩處
            4.電極脫水后儀表指示有數值
            A.電極透水或電纜絕緣性能下降
            B.電極間有異物
            C.接線端子受潮
            A.更換新電極

            B.C.從查明原因針對處理
            5.報警有訊響繼電器不動作
            A.繼電器有響動,觸點不切換位置
            B.繼電器無響動
            A.更換繼電器(觸點燒蝕)
            B.電路故障,返廠維修